The NRC autocollimator calibration facility

Download
  1. Get@NRC: The NRC autocollimator calibration facility (Opens in a new window)
DOIResolve DOI: http://doi.org/10.1088/0026-1394/50/5/433
AuthorSearch for:
TypeArticle
Journal titleMetrologia
ISSN0026-1394
Volume50
Issue5
Pages433440; # of pages: 8
SubjectAutocollimators; Rotary encoder; Rotary tables; Scale Factor; Standard uncertainty; Plant shutdowns; Calibration
AbstractAn autocollimator calibration facility has been developed based upon both a precision rotary table and a sine-bar. The traceability of the system is through sub-division of the circle and hence to the SI unit of angle, and does not require traceability to the definition of the metre. The rotary encoder that measures the angle of the rotary table is calibrated through circle closure techniques to a combined standard uncertainty of 0.013 arcsec at 5° intervals. The rotation of the sine-bar is measured by two interferometers, and together with the rotary encoder, enables calibration of the scale factor of an autocollimator to a combined standard uncertainty of 1.8 × 10 -5. © 2013 BIPM & IOP Publishing Ltd.
Publication date
LanguageEnglish
AffiliationNational Research Council Canada (NRC-CNRC)
Peer reviewedYes
NPARC number21270634
Export citationExport as RIS
Report a correctionReport a correction
Record identifier0bf8bb9b-b5f4-4274-97c2-bc6a7c000470