An analysis of inconsistencies between ITS-90 interpolations above 0.01 °c

Download
  1. Get@NRC: An analysis of inconsistencies between ITS-90 interpolations above 0.01 °c (Opens in a new window)
DOIResolve DOI: http://doi.org/10.1007/s10765-010-0891-9
AuthorSearch for: ; Search for: ; Search for: ; Search for: ; Search for: ; Search for:
TypeArticle
Journal titleInternational Journal of Thermophysics
ISSN0195-928X
Volume32
Issue1-Feb
Pages6885; # of pages: 18
SubjectInconsistency; ITS-90; Overlap region; Standard platinum resistance thermometers; Platinum; Thermometers; Interpolation
AbstractInconsistencies in the ITS-90 interpolations above 0.01 °C have been investigated. Thirty-five standard platinum resistance thermometers are used to determine the magnitude of inconsistencies in this range with the exception of the overlap region of the water-zinc and water-aluminum sub-ranges. The results show inconsistencies that are less 4.1 mK. Simpler representations of ten of the fifteen inconsistency functions above 0.01 °C are presented by means of factoring and extremum analysis. © 2010 Springer Science+Business Media, LLC.