Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the NRC, Canada and the BIPM in low-energy x-rays

Download
  1. (PDF, 366 KB)
  2. Get@NRC: Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the NRC, Canada and the BIPM in low-energy x-rays (Opens in a new window)
DOIResolve DOI: http://doi.org/10.1088/0026-1394/48/1A/06002
AuthorSearch for: ; Search for:
TypeArticle
Journal titleMetrologia
Volume48
IssueTechnical Supplement 2011
Pages112; # of pages: 12
AbstractA key comparison has been made between the air-kerma standards of the NRC and the BIPM in the low-energy x ray range. The results show the standards to be in agreement at the level of the combined standard uncertainty of 2.8 parts in 103. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Publication date
LanguageEnglish
AffiliationNational Research Council Canada; NRC Institute for National Measurement Standards
Peer reviewedYes
NPARC number16774803
Export citationExport as RIS
Report a correctionReport a correction
Record identifier2e39e274-fb3d-40f1-a27c-a3b937e33472
Record created2011-02-01
Record modified2016-05-09
Bookmark and share
  • Share this page with Facebook (Opens in a new window)
  • Share this page with Twitter (Opens in a new window)
  • Share this page with Google+ (Opens in a new window)
  • Share this page with Delicious (Opens in a new window)