Experimental determination of megavoltage electron beam k'R50 factors using water calorimetry

Download
  1. Get@NRC: Experimental determination of megavoltage electron beam k'R50 factors using water calorimetry (Opens in a new window)
DOIResolve DOI: http://doi.org/10.1016/S0167-8140(15)30923-3
AuthorSearch for: ; Search for: ; Search for: ; Search for: ; Search for: ; Search for:
TypeArticle
Proceedings titleRadiotherapy and Oncology
ConferenceESTRO 33, 4-8 April 2014, Vienna, Austria
ISSN0167-8140
Volume111
Article numberSupplement 1
PagesS62
AbstractIn this work, electron quality conversion factors, K'R50, for both a cylindrical ion chamber and a parallel plate ion chamber are measured for electron beam energies of 6, 9, 12, 16, and 20 MeV, using a water calorimeter. The photon-electron conversion factor, k ecal, for the cylindrical chamber is derived from calorimetry measurements in 60 Co. The purpose of this work is to experimentally verify the accuracy of values included in current dosimetry protocols.
Publication date
LanguageEnglish
AffiliationMeasurement Science and Standards; National Research Council Canada
Peer reviewedYes
NPARC number21275151
Export citationExport as RIS
Report a correctionReport a correction
Record identifier77d48b97-77e8-4048-a44d-7026e783b030
Record created2015-05-20
Record modified2016-05-09
Bookmark and share
  • Share this page with Facebook (Opens in a new window)
  • Share this page with Twitter (Opens in a new window)
  • Share this page with Google+ (Opens in a new window)