Cobalt–carbon eutectic fixed point for contact thermometry

Download
  1. (PDF, 2 MB)
  2. Get@NRC: Cobalt–carbon eutectic fixed point for contact thermometry (Opens in a new window)
DOIResolve DOI: http://doi.org/10.1007/s10765-010-0907-5
AuthorSearch for: ; Search for: ; Search for: ; Search for:
TypeArticle
Journal titleInternational Journal of Thermophysics
Volume32
Issue1-2
Pages453462; # of pages: 9
SubjectEutectics; High-temperature fixed points; Pd–C; Pt/Pd Thermocouples
AbstractTwo Co–C eutectic fixed points were constructed for thermocouple calibration. The eutectic fixed points were measured with a Pt/Pd thermocouple calibrated at the freezing temperatures of Sn, Zn, Al, Ag, and Au. A temperature of (1323.99 ± 0.52)◦C(k = 2) was determined via thismethod. The cell design allowed filling to be accomplished in a single step. Each cellwas held above 1300 ◦C for at least 42 h and was subjected to at least 20 melt/freeze cycles with no mechanical failure occurring.abstract