Calibration of 10∶−1 inductive voltage dividers by Thompson's method using an electronically-enhanced three-stage calibrating transformer

DOIResolve DOI:
AuthorSearch for:
Conference2012 Conference on Precision Electromagnetic Measurements (CPEM 2012), 2012-07-1 - 2012-07-6, Washington, DC, USA
Pages500501; # of pages: 2
SubjectInductive voltage dividers; calibration; ratio transformers; voltage comparator
AbstractThis paper describes the calibration of 10:-1 inductive voltage dividers at power frequencies using Thompson's method. It also describes an electronically-enhanced three-stage 10:-1 inductive voltage divider (IVD) with an additional, secondary winding, used as a calibrating transformer. The correction circuitry was designed to further reduce the calibrating transformer magnetic error and is isolated from the input and output voltages. To define open-circuit voltage at IVD and calibrating transformer outputs, a BNC connector was modified with an additional electrode and used as part of the special socket with a switch. The calibration system was applied to the calibration of two two-stage IVDs.