Espectrómetro de transformada de fourier implementado mediante guías de onda en espiral con base de silicio

Download
  1. Get@NRC: Espectrómetro de transformada de fourier implementado mediante guías de onda en espiral con base de silicio (Opens in a new window)
DOIResolve DOI: http://doi.org/10.7149/OPA.46.2.99
AuthorSearch for: ; Search for: ; Search for: ; Search for: ; Search for: ; Search for: ; Search for: ; Search for: ; Search for: ; Search for:
TypeArticle
Journal titleOptica Pura y Aplicada
ISSN0030-3917
Volume46
Issue2
Pages99103; # of pages: 5
AbstractIn this work we present the design, fabrication and characterization of a Fourier-Transform micro-spectrometer implemented with planar waveguides. The device comprises an array of 32 Mach-Zehnder interferometers with increasing optical path differences between the arms of the interferometer, reaching a resolution of 0.1 nm in a free spectral range of 1.6 nm. Silicon waveguides in a spiral structure have been used in the arms of the interferometer, reducing the footprint of the device below 12 mm2.
Publication date
AffiliationNRC Institute for Microstructural Sciences; National Research Council Canada