TY - CONF ID - 20955690 AU - Filipski, Piotr S. T1 - Investigation of systematic errors of an AC Jospehson Voltage Standard C8 - 2012 Conference on Precision Electromagnetic Measurements T3 - 2012 Conference on Precision Electromagnetic Measurements (CPEM 2012), 1-6 July 2012, Washington, DC, USA PB - IEEE SP - 36 EP - 37 PY - 2012 KW - Josephson effect KW - calibration KW - digital-analog conversion KW - superconducting arrays KW - voltage measurement KW - signal synthesis N2 - This paper presents results of experiments quantifying two systematic errors associated with the calibrations by use of an AC Josephson Voltage Standard: the change in the voltage leads error with the change in the level of probe immersion in the liquid helium, and the effect of the internal inductance of a Josephson array. The measurements were performed at frequencies 100 kHz and above, however, they can be used in evaluating an error budget at lower frequencies. DO - http://dx.doi.org/10.1109/CPEM.2012.6250646 U3 - NRC Institute for National Measurement Standards U3 - Institut des étalons nationaux de mesure du CNRC (IENM-INMS) U3 - National Research Council Canada U3 - Conseil national de recherches Canada (CNRC-NRC) SN - 0589-1485 SN - 978-1-4673-0439-9 ER -