Government of Canada
Symbol of the Government of Canada
 
 
 
 

Investigation of systematic errors of an AC Jospehson Voltage Standard

 
 
Affiliation:
NRC Institute for National Measurement Standards; National Research Council Canada
Language:
English
Type:
Conference publication
Conference:
2012 Conference on Precision Electromagnetic Measurements (CPEM 2012), 1-6 July 2012, Washington, DC, USA
Proceedings
Title:
2012 Conference on Precision Electromagnetic Measurements
ISSN:
0589-1485
Date:
2012
Pages :
36-37
NPArC #:
20955690
Keywords:
Josephson effect; calibration; digital-analog conversion; superconducting arrays; voltage measurement; signal synthesis
Abstract:
This paper presents results of experiments quantifying two systematic errors associated with the calibrations by use of an AC Josephson Voltage Standard: the change in the voltage leads error with the change in the level of probe immersion in the liquid helium, and the effect of the internal inductance of a Josephson array. The measurements were performed at frequencies 100 kHz and above, however, they can be used in evaluating an error budget at lower frequencies.
 